A system for measuring elastic wave velocity under high pressure and high temperature using a combination of ultrasonic measurement and the multi-anvil apparatus at SPring-8

An experimental system to measure the elastic wave velocities of hot-pressed polycrystalline samples at high pressure and high temperature has been installed at SPring-8. It uses a combination of the ultrasonic pulse-echo-overlap method and Kawai-type multi-anvil apparatus (SPEED-1500). X-ray radiog...

Ausführliche Beschreibung

Bibliographische Detailangaben
Veröffentlicht in:Journal of synchrotron radiation. - 1994. - 16(2009), Pt 6 vom: 18. Nov., Seite 762-8
1. Verfasser: Higo, Yuji (VerfasserIn)
Weitere Verfasser: Kono, Yoshio, Inoue, Toru, Irifune, Tetsuo, Funakoshi, Ken-ichi
Format: Online-Aufsatz
Sprache:English
Veröffentlicht: 2009
Zugriff auf das übergeordnete Werk:Journal of synchrotron radiation
Schlagworte:Journal Article Research Support, Non-U.S. Gov't
Beschreibung
Zusammenfassung:An experimental system to measure the elastic wave velocities of hot-pressed polycrystalline samples at high pressure and high temperature has been installed at SPring-8. It uses a combination of the ultrasonic pulse-echo-overlap method and Kawai-type multi-anvil apparatus (SPEED-1500). X-ray radiographic imaging enables the sample length to be determined at high pressure and high temperature, which is indispensable for precise determination of elastic wave velocity. Precise measurements of the elastic wave velocities of various minerals have been determined at pressures up to 19 GPa and temperatures up to 1673 K. The experimental technique provides the precise elastic wave velocities of various materials corresponding to those in the Earth's mantle
Beschreibung:Date Completed 05.01.2010
Date Revised 21.10.2009
published: Print-Electronic
Citation Status PubMed-not-MEDLINE
ISSN:1600-5775
DOI:10.1107/S0909049509034980