An ultrasonic imaging speckle-suppression and contrast-enhancement technique by means of frequency compounding and coded excitation

A method for improving the contrast resolution of B-mode images is proposed by combining the speckle-reduction technique of frequency compounding (FC) and the coded excitation and pulse-compression technique called resolution enhancement compression (REC). FC suppresses speckle but at the expense of...

Description complète

Détails bibliographiques
Publié dans:IEEE transactions on ultrasonics, ferroelectrics, and frequency control. - 1986. - 56(2009), 7 vom: 15. Juli, Seite 1327-39
Auteur principal: Sanchez, Jose R (Auteur)
Autres auteurs: Oelze, Michael L
Format: Article en ligne
Langue:English
Publié: 2009
Accès à la collection:IEEE transactions on ultrasonics, ferroelectrics, and frequency control
Sujets:Journal Article Research Support, N.I.H., Extramural
LEADER 01000caa a22002652c 4500
001 NLM189733489
003 DE-627
005 20250210134851.0
007 cr uuu---uuuuu
008 231223s2009 xx |||||o 00| ||eng c
024 7 |a 10.1109/TUFFC.2009.1189  |2 doi 
028 5 2 |a pubmed25n0633.xml 
035 |a (DE-627)NLM189733489 
035 |a (NLM)19574144 
040 |a DE-627  |b ger  |c DE-627  |e rakwb 
041 |a eng 
100 1 |a Sanchez, Jose R  |e verfasserin  |4 aut 
245 1 3 |a An ultrasonic imaging speckle-suppression and contrast-enhancement technique by means of frequency compounding and coded excitation 
264 1 |c 2009 
336 |a Text  |b txt  |2 rdacontent 
337 |a ƒaComputermedien  |b c  |2 rdamedia 
338 |a ƒa Online-Ressource  |b cr  |2 rdacarrier 
500 |a Date Completed 22.10.2009 
500 |a Date Revised 03.07.2009 
500 |a published: Print 
500 |a Citation Status MEDLINE 
520 |a A method for improving the contrast resolution of B-mode images is proposed by combining the speckle-reduction technique of frequency compounding (FC) and the coded excitation and pulse-compression technique called resolution enhancement compression (REC). FC suppresses speckle but at the expense of a reduction in axial resolution. Using REC, the axial resolution and bandwidth of the imaging system was doubled. Therefore, by combining REC with FC (REC-FC), the tradeoff between axial resolution and contrast enhancement was extended significantly. Simulations and experimental measurements were conducted with a single-element transducer (f/2.66) having a center frequency of 2.25 MHz and a -3-dB bandwidth of 50%. Simulations and measurements of hyperechoic (+6 dB) tissue-mimicking targets were imaged. Four FC cases were evaluated: full-, half-, third-, and fourth-width of the true impulse response bandwidth. The image quality metrics used to compare REC-FC to conventional pulsing (CP) and CP-FC were contrast-to-noise ratio (CNR), speckle signal-to-noise ratio, histogram pixel intensity, and lesion signal-to-noise ratio. Increases in CNR of 121%, 231%, 302%, and 391% were obtained in experiments when comparing REC-FC for the full-, half-, third-, and fourth-width cases to CP. Furthermore, smaller increases in CNR of 112%, 233%, and 309% were obtained in experiments when comparing CP-FC for the half-, third-, and fourth-width cases to CP. Improved lesion detectability was observed by using REC-FC 
650 4 |a Journal Article 
650 4 |a Research Support, N.I.H., Extramural 
700 1 |a Oelze, Michael L  |e verfasserin  |4 aut 
773 0 8 |i Enthalten in  |t IEEE transactions on ultrasonics, ferroelectrics, and frequency control  |d 1986  |g 56(2009), 7 vom: 15. Juli, Seite 1327-39  |w (DE-627)NLM098181017  |x 1525-8955  |7 nnas 
773 1 8 |g volume:56  |g year:2009  |g number:7  |g day:15  |g month:07  |g pages:1327-39 
856 4 0 |u http://dx.doi.org/10.1109/TUFFC.2009.1189  |3 Volltext 
912 |a GBV_USEFLAG_A 
912 |a SYSFLAG_A 
912 |a GBV_NLM 
912 |a GBV_ILN_22 
912 |a GBV_ILN_24 
912 |a GBV_ILN_350 
951 |a AR 
952 |d 56  |j 2009  |e 7  |b 15  |c 07  |h 1327-39