Broadband ultrasound attenuation measurement of long bone using peak frequency of the echoes
The traditional peak frequency formulation requires the knowledge of the signal wavelet. We improved and applied the method to estimate attenuation for homogeneous silicon rubber and bovine cortical bone without recourse to the wavelet assumption. The estimated values for rubber and bone samples are...
Veröffentlicht in: | IEEE transactions on ultrasonics, ferroelectrics, and frequency control. - 1986. - 56(2009), 2 vom: 01. Feb., Seite 396-9 |
---|---|
1. Verfasser: | |
Weitere Verfasser: | , , |
Format: | Online-Aufsatz |
Sprache: | English |
Veröffentlicht: |
2009
|
Zugriff auf das übergeordnete Werk: | IEEE transactions on ultrasonics, ferroelectrics, and frequency control |
Schlagworte: | Letter Research Support, Non-U.S. Gov't Silicone Elastomers |
Zusammenfassung: | The traditional peak frequency formulation requires the knowledge of the signal wavelet. We improved and applied the method to estimate attenuation for homogeneous silicon rubber and bovine cortical bone without recourse to the wavelet assumption. The estimated values for rubber and bone samples are 6.59 +/- 0.28 dB/MHz/cm and 4.59 +/- 1.09 dB/MHz/cm, respectively, as compared with 6.33 +/- 0.19 dB/ MHz/cm and 5.00 +/- 1.10 dB/MHz/cm, respectively, by the spectral ratio method |
---|---|
Beschreibung: | Date Completed 17.08.2009 Date Revised 25.11.2016 published: Print Citation Status MEDLINE |
ISSN: | 1525-8955 |
DOI: | 10.1109/TUFFC.2009.1049 |