A curved image-plate detector system for high-resolution synchrotron X-ray diffraction

The developed curved image plate (CIP) is a one-dimensional detector which simultaneously records high-resolution X-ray diffraction (XRD) patterns over a 38.7 degrees 2theta range. In addition, an on-site reader enables rapid extraction, transfer and storage of X-ray intensity information in </=3...

Ausführliche Beschreibung

Bibliographische Detailangaben
Veröffentlicht in:Journal of synchrotron radiation. - 1994. - 16(2009), Pt 2 vom: 13. März, Seite 273-82
1. Verfasser: Sarin, P (VerfasserIn)
Weitere Verfasser: Haggerty, R P, Yoon, W, Knapp, M, Berghaeuser, A, Zschack, P, Karapetrova, E, Yang, N, Kriven, W M
Format: Online-Aufsatz
Sprache:English
Veröffentlicht: 2009
Zugriff auf das übergeordnete Werk:Journal of synchrotron radiation
Schlagworte:Journal Article Research Support, Non-U.S. Gov't Research Support, U.S. Gov't, Non-P.H.S.
LEADER 01000caa a22002652 4500
001 NLM186653255
003 DE-627
005 20250210052622.0
007 cr uuu---uuuuu
008 231223s2009 xx |||||o 00| ||eng c
024 7 |a 10.1107/S0909049509001265  |2 doi 
028 5 2 |a pubmed25n0622.xml 
035 |a (DE-627)NLM186653255 
035 |a (NLM)19240340 
040 |a DE-627  |b ger  |c DE-627  |e rakwb 
041 |a eng 
100 1 |a Sarin, P  |e verfasserin  |4 aut 
245 1 2 |a A curved image-plate detector system for high-resolution synchrotron X-ray diffraction 
264 1 |c 2009 
336 |a Text  |b txt  |2 rdacontent 
337 |a ƒaComputermedien  |b c  |2 rdamedia 
338 |a ƒa Online-Ressource  |b cr  |2 rdacarrier 
500 |a Date Completed 28.04.2009 
500 |a Date Revised 25.02.2009 
500 |a published: Print-Electronic 
500 |a Citation Status MEDLINE 
520 |a The developed curved image plate (CIP) is a one-dimensional detector which simultaneously records high-resolution X-ray diffraction (XRD) patterns over a 38.7 degrees 2theta range. In addition, an on-site reader enables rapid extraction, transfer and storage of X-ray intensity information in </=30 s, and further qualifies this detector to study kinetic processes in materials science. The CIP detector can detect and store X-ray intensity information linearly proportional to the incident photon flux over a dynamical range of about five orders of magnitude. The linearity and uniformity of the CIP detector response is not compromised in the unsaturated regions of the image plate, regardless of saturation in another region. The speed of XRD data acquisition together with excellent resolution afforded by the CIP detector is unique and opens up wide possibilities in materials research accessible through X-ray diffraction. This article presents details of the basic features, operation and performance of the CIP detector along with some examples of applications, including high-temperature XRD 
650 4 |a Journal Article 
650 4 |a Research Support, Non-U.S. Gov't 
650 4 |a Research Support, U.S. Gov't, Non-P.H.S. 
700 1 |a Haggerty, R P  |e verfasserin  |4 aut 
700 1 |a Yoon, W  |e verfasserin  |4 aut 
700 1 |a Knapp, M  |e verfasserin  |4 aut 
700 1 |a Berghaeuser, A  |e verfasserin  |4 aut 
700 1 |a Zschack, P  |e verfasserin  |4 aut 
700 1 |a Karapetrova, E  |e verfasserin  |4 aut 
700 1 |a Yang, N  |e verfasserin  |4 aut 
700 1 |a Kriven, W M  |e verfasserin  |4 aut 
773 0 8 |i Enthalten in  |t Journal of synchrotron radiation  |d 1994  |g 16(2009), Pt 2 vom: 13. März, Seite 273-82  |w (DE-627)NLM09824129X  |x 1600-5775  |7 nnns 
773 1 8 |g volume:16  |g year:2009  |g number:Pt 2  |g day:13  |g month:03  |g pages:273-82 
856 4 0 |u http://dx.doi.org/10.1107/S0909049509001265  |3 Volltext 
912 |a GBV_USEFLAG_A 
912 |a SYSFLAG_A 
912 |a GBV_NLM 
912 |a GBV_ILN_40 
912 |a GBV_ILN_350 
912 |a GBV_ILN_2005 
951 |a AR 
952 |d 16  |j 2009  |e Pt 2  |b 13  |c 03  |h 273-82