Design and fabrication of PZN-7%PT single crystal high frequency angled needle ultrasound transducers

A high-frequency angled needle ultrasound transducer with an aperture size of 0.4 x 0.56 mm2 was fabricated using a lead zinc niobate-lead titanate (PZN- 7%PT) single crystal as the active piezoelectric material. The single crystal was bonded to a conductive silver particle matching layer and a cond...

Ausführliche Beschreibung

Bibliographische Detailangaben
Veröffentlicht in:IEEE transactions on ultrasonics, ferroelectrics, and frequency control. - 1986. - 55(2008), 6 vom: 10., Seite 1394-9
1. Verfasser: Zhou, Qifa (VerfasserIn)
Weitere Verfasser: Wu, Dawei, Jin, Jing, Hu, Chang-hong, Xu, Xiaochen, Williams, Jay, Cannata, Jonathan M, Lim, Leongchew, Shung, K Kirk
Format: Online-Aufsatz
Sprache:English
Veröffentlicht: 2008
Zugriff auf das übergeordnete Werk:IEEE transactions on ultrasonics, ferroelectrics, and frequency control
Schlagworte:Letter Research Support, N.I.H., Extramural lead zinc niobate titanate Niobium 05175J654G Lead 2P299V784P Titanium D1JT611TNE Zinc J41CSQ7QDS
Beschreibung
Zusammenfassung:A high-frequency angled needle ultrasound transducer with an aperture size of 0.4 x 0.56 mm2 was fabricated using a lead zinc niobate-lead titanate (PZN- 7%PT) single crystal as the active piezoelectric material. The single crystal was bonded to a conductive silver particle matching layer and a conductive epoxy backing material through direct contact curing. A parylene outer matching layer was formed by vapor deposition. Angled needle probe configuration was achieved by dicing at 45 degrees to the single crystal poling direction to satisfy a clinical request for blood flow measurement in the posterior portion of the eye. The electrical impedance magnitude and phase of the transducer were 42 Omega and -63 degrees , respectively. The measured center frequency and the fractional bandwidth at -6 dB were 43 MHz and 45%, respectively. The two-way insertion loss was approximately 17 dB. Wire phantom imaging using fabricated PZN-7%PT single crystal transducers was obtained and spatial resolutions were assessed
Beschreibung:Date Completed 02.09.2008
Date Revised 29.05.2025
published: Print
Citation Status MEDLINE
ISSN:1525-8955