Evaluation of two methods to minimize the anticlastic curvature of cylindrical bent crystals

A scanning dynamically focusing sagittal X-ray monochromator accepting 3 mrad x 0.1 mrad from a 6 T wiggler has been designed for XAFS measurements. In the energy range 4.1-12.4 keV, the slope error of the second cylindrical crystal caused by anticlastic bending must be less than 1/5 of the Darwin w...

Ausführliche Beschreibung

Bibliographische Detailangaben
Veröffentlicht in:Journal of synchrotron radiation. - 1994. - 15(2008), Pt 2 vom: 07. März, Seite 140-3
1. Verfasser: Feng, Liangjie (VerfasserIn)
Weitere Verfasser: Kang, Le, Li, Zhongliang, Zhao, Feixun, Xu, Chaoyin
Format: Online-Aufsatz
Sprache:English
Veröffentlicht: 2008
Zugriff auf das übergeordnete Werk:Journal of synchrotron radiation
Schlagworte:Journal Article
LEADER 01000caa a22002652c 4500
001 NLM177828633
003 DE-627
005 20250209062613.0
007 cr uuu---uuuuu
008 231223s2008 xx |||||o 00| ||eng c
024 7 |a 10.1107/S0909049507066368  |2 doi 
028 5 2 |a pubmed25n0593.xml 
035 |a (DE-627)NLM177828633 
035 |a (NLM)18296779 
040 |a DE-627  |b ger  |c DE-627  |e rakwb 
041 |a eng 
100 1 |a Feng, Liangjie  |e verfasserin  |4 aut 
245 1 0 |a Evaluation of two methods to minimize the anticlastic curvature of cylindrical bent crystals 
264 1 |c 2008 
336 |a Text  |b txt  |2 rdacontent 
337 |a ƒaComputermedien  |b c  |2 rdamedia 
338 |a ƒa Online-Ressource  |b cr  |2 rdacarrier 
500 |a Date Completed 19.05.2008 
500 |a Date Revised 25.02.2008 
500 |a published: Print-Electronic 
500 |a Citation Status PubMed-not-MEDLINE 
520 |a A scanning dynamically focusing sagittal X-ray monochromator accepting 3 mrad x 0.1 mrad from a 6 T wiggler has been designed for XAFS measurements. In the energy range 4.1-12.4 keV, the slope error of the second cylindrical crystal caused by anticlastic bending must be less than 1/5 of the Darwin width of the crystal or the photon flux will decrease drastically. Two methods to minimize the anticlastic bending are proposed. Thin crystals with stiffening ribs and thin crystals with an aspect ratio equal to the ;golden value' are evaluated by finite-element analysis and by long-trace-profiler characterization. Both approaches are satisfactory, but the ;golden value' approach is preferred in this case for the second crystal of the new monochromator not only because it is easy to manufacture but also because the surface is smoother than the ribbed crystal 
650 4 |a Journal Article 
700 1 |a Kang, Le  |e verfasserin  |4 aut 
700 1 |a Li, Zhongliang  |e verfasserin  |4 aut 
700 1 |a Zhao, Feixun  |e verfasserin  |4 aut 
700 1 |a Xu, Chaoyin  |e verfasserin  |4 aut 
773 0 8 |i Enthalten in  |t Journal of synchrotron radiation  |d 1994  |g 15(2008), Pt 2 vom: 07. März, Seite 140-3  |w (DE-627)NLM09824129X  |x 1600-5775  |7 nnas 
773 1 8 |g volume:15  |g year:2008  |g number:Pt 2  |g day:07  |g month:03  |g pages:140-3 
856 4 0 |u http://dx.doi.org/10.1107/S0909049507066368  |3 Volltext 
912 |a GBV_USEFLAG_A 
912 |a SYSFLAG_A 
912 |a GBV_NLM 
912 |a GBV_ILN_40 
912 |a GBV_ILN_350 
912 |a GBV_ILN_2005 
951 |a AR 
952 |d 15  |j 2008  |e Pt 2  |b 07  |c 03  |h 140-3