A computerized system for measuring the acoustic output from diagnostic ultrasound equipment

The measurement arrangement, which complies with the requirements of the US Food and Drug Administration, consists of a positioning system with a full range of degrees-of-freedom and a digital oscilloscope, both under complete computer control. The acoustic pressure-time waveform is recorded using m...

Ausführliche Beschreibung

Bibliographische Detailangaben
Veröffentlicht in:IEEE transactions on ultrasonics, ferroelectrics, and frequency control. - 1986. - 35(1988), 2 vom: 15., Seite 102-9
1. Verfasser: Schafer, M E (VerfasserIn)
Weitere Verfasser: Lewin, P A
Format: Aufsatz
Sprache:English
Veröffentlicht: 1988
Zugriff auf das übergeordnete Werk:IEEE transactions on ultrasonics, ferroelectrics, and frequency control
Schlagworte:Journal Article
LEADER 01000naa a22002652 4500
001 NLM177765003
003 DE-627
005 20231223150936.0
007 tu
008 231223s1988 xx ||||| 00| ||eng c
028 5 2 |a pubmed24n0593.xml 
035 |a (DE-627)NLM177765003 
035 |a (NLM)18290136 
040 |a DE-627  |b ger  |c DE-627  |e rakwb 
041 |a eng 
100 1 |a Schafer, M E  |e verfasserin  |4 aut 
245 1 2 |a A computerized system for measuring the acoustic output from diagnostic ultrasound equipment 
264 1 |c 1988 
336 |a Text  |b txt  |2 rdacontent 
337 |a ohne Hilfsmittel zu benutzen  |b n  |2 rdamedia 
338 |a Band  |b nc  |2 rdacarrier 
500 |a Date Completed 02.10.2012 
500 |a Date Revised 21.02.2008 
500 |a published: Print 
500 |a Citation Status PubMed-not-MEDLINE 
520 |a The measurement arrangement, which complies with the requirements of the US Food and Drug Administration, consists of a positioning system with a full range of degrees-of-freedom and a digital oscilloscope, both under complete computer control. The acoustic pressure-time waveform is recorded using membrane-type and needle-type polyvinylidene fluoride (PVDF) hydrophone probes. The overall bandwidth of the system depends on the hydrophone probe used and can range up to 100 MHz. A complete description of the system and the measurement procedures is given, along with a brief discussion of the various factors which affect measurement uncertainty. The largest overall uncertainty of the same associated with acoustic intensity measurements was determined to be no greater than 20% for I(sppa) and 25% for I(spta) (spatial peak pulse average intensity and spatial-peak temporal-average intensity, respectively). Other applications of the system include transducer characterization and research work in ultrasound dosimetry and bioeffects 
650 4 |a Journal Article 
700 1 |a Lewin, P A  |e verfasserin  |4 aut 
773 0 8 |i Enthalten in  |t IEEE transactions on ultrasonics, ferroelectrics, and frequency control  |d 1986  |g 35(1988), 2 vom: 15., Seite 102-9  |w (DE-627)NLM098181017  |x 1525-8955  |7 nnns 
773 1 8 |g volume:35  |g year:1988  |g number:2  |g day:15  |g pages:102-9 
912 |a GBV_USEFLAG_A 
912 |a SYSFLAG_A 
912 |a GBV_NLM 
912 |a GBV_ILN_22 
912 |a GBV_ILN_24 
912 |a GBV_ILN_350 
951 |a AR 
952 |d 35  |j 1988  |e 2  |b 15  |h 102-9