A real-time ultrasound time-domain correlation blood flowmeter. II. Performance and experimental verification

For Part I, see ibid., vol.40, no.6, p.768-775 (1993). The performance of a real-time ultrasound time-domain correlation (UTDC) flowmeter employing a high-speed residue number system correlator under various flow conditions for different correlation parameter settings is evaluated. Previous work has...

Ausführliche Beschreibung

Bibliographische Detailangaben
Veröffentlicht in:IEEE transactions on ultrasonics, ferroelectrics, and frequency control. - 1986. - 40(1993), 6 vom: 15., Seite 776-85
1. Verfasser: Hein, I A (VerfasserIn)
Weitere Verfasser: O'Brien, W R
Format: Aufsatz
Sprache:English
Veröffentlicht: 1993
Zugriff auf das übergeordnete Werk:IEEE transactions on ultrasonics, ferroelectrics, and frequency control
Schlagworte:Journal Article
LEADER 01000caa a22002652 4500
001 NLM177507128
003 DE-627
005 20250209050826.0
007 tu
008 231223s1993 xx ||||| 00| ||eng c
028 5 2 |a pubmed25n0592.xml 
035 |a (DE-627)NLM177507128 
035 |a (NLM)18263246 
040 |a DE-627  |b ger  |c DE-627  |e rakwb 
041 |a eng 
100 1 |a Hein, I A  |e verfasserin  |4 aut 
245 1 2 |a A real-time ultrasound time-domain correlation blood flowmeter. II. Performance and experimental verification 
264 1 |c 1993 
336 |a Text  |b txt  |2 rdacontent 
337 |a ohne Hilfsmittel zu benutzen  |b n  |2 rdamedia 
338 |a Band  |b nc  |2 rdacarrier 
500 |a Date Completed 02.10.2012 
500 |a Date Revised 11.02.2008 
500 |a published: Print 
500 |a Citation Status PubMed-not-MEDLINE 
520 |a For Part I, see ibid., vol.40, no.6, p.768-775 (1993). The performance of a real-time ultrasound time-domain correlation (UTDC) flowmeter employing a high-speed residue number system correlator under various flow conditions for different correlation parameter settings is evaluated. Previous work has incorporated a weighted-averaging scheme to estimate the flow velocity over a wide number of echo signal spacings. The present evaluation indicates that a linear averaging of 50 adjacent echo signal pairs produces equivalent accuracy and precision at real-time processing speed as compared to previous non-real time systems incorporating weighted averaging 
650 4 |a Journal Article 
700 1 |a O'Brien, W R  |e verfasserin  |4 aut 
773 0 8 |i Enthalten in  |t IEEE transactions on ultrasonics, ferroelectrics, and frequency control  |d 1986  |g 40(1993), 6 vom: 15., Seite 776-85  |w (DE-627)NLM098181017  |x 1525-8955  |7 nnns 
773 1 8 |g volume:40  |g year:1993  |g number:6  |g day:15  |g pages:776-85 
912 |a GBV_USEFLAG_A 
912 |a SYSFLAG_A 
912 |a GBV_NLM 
912 |a GBV_ILN_22 
912 |a GBV_ILN_24 
912 |a GBV_ILN_350 
951 |a AR 
952 |d 40  |j 1993  |e 6  |b 15  |h 776-85