Invariance of the Doppler bandwidth with range cell size above a critical beam-to-flow angle

For a sound beam impinging on a blood vessel, with a range cell much smaller than the vessel diameter, it is known that the breadth of the echo Doppler spectrum is proportional to the velocity of the flow through the range cell. As the range cell is lengthened to include a greater range of velocitie...

Ausführliche Beschreibung

Bibliographische Detailangaben
Veröffentlicht in:IEEE transactions on ultrasonics, ferroelectrics, and frequency control. - 1986. - 40(1993), 4 vom: 15., Seite 381-6
1. Verfasser: Tortoli, P (VerfasserIn)
Weitere Verfasser: Guidi, G, Newhouse, V L
Format: Aufsatz
Sprache:English
Veröffentlicht: 1993
Zugriff auf das übergeordnete Werk:IEEE transactions on ultrasonics, ferroelectrics, and frequency control
Schlagworte:Journal Article
Beschreibung
Zusammenfassung:For a sound beam impinging on a blood vessel, with a range cell much smaller than the vessel diameter, it is known that the breadth of the echo Doppler spectrum is proportional to the velocity of the flow through the range cell. As the range cell is lengthened to include a greater range of velocities, the spectrum is expected to widen proportionately. It is shown theoretically, and confirmed experimentally, that if the beam-to-flow angle is greater than a critical value, the Doppler spectrum bandwidth is independent of the length of the range cell, and depends only on the maximum velocity encompassed by it. This happens because for angles greater than the critical, the narrow spectra produced by lower velocity flows near the vessel walls are contained within the broader spectrum produced by the higher speed flow near the vessel axis. The critical angle is the angle at which the flow axis is normal to one of the beam edges
Beschreibung:Date Completed 02.10.2012
Date Revised 11.02.2008
published: Print
Citation Status PubMed-not-MEDLINE
ISSN:1525-8955