Impact of propagation through an aberrating medium on the linear effective apodization of a nonlinearly generated second harmonic field

Techniques based on the nonlinearly generated second harmonic signal (tissue harmonic imaging) have rapidly supplanted linear (fundamental) imaging methods as the standard in two-dimensional echocardiography. Enhancements to the compactness of the nonlinearly generated second harmonic (2f) field com...

Ausführliche Beschreibung

Bibliographische Detailangaben
Veröffentlicht in:IEEE transactions on ultrasonics, ferroelectrics, and frequency control. - 1999. - 53(2006), 7 vom: 29. Juli, Seite 1260-8
1. Verfasser: Wallace, Kirk D (VerfasserIn)
Weitere Verfasser: Holland, Mark R, Robinson, Brent S, Fedewa, Russell J, Lloyd, Christopher W, Miller, James G
Format: Aufsatz
Sprache:English
Veröffentlicht: 2006
Zugriff auf das übergeordnete Werk:IEEE transactions on ultrasonics, ferroelectrics, and frequency control
Schlagworte:Journal Article Research Support, N.I.H., Extramural
Beschreibung
Zusammenfassung:Techniques based on the nonlinearly generated second harmonic signal (tissue harmonic imaging) have rapidly supplanted linear (fundamental) imaging methods as the standard in two-dimensional echocardiography. Enhancements to the compactness of the nonlinearly generated second harmonic (2f) field component with respect to the fundamental (1f) field component are widely considered to be among the factors contributing to the observed image quality improvements. The objective of this study was to measure the impact of phase and amplitude aberrations resulting from propagation through an inhomogeneous tissue, on the beamwidths associated with: the fundamental (1f); the nonlinearly generated second harmonic (2f); and the linearly propagated, effective apodization signal at the same (21) frequency. Modifications to the transmit characteristics of a phased-array imaging system were validated with hydrophone measurements. Results demonstrate that the characteristics of the diffraction pattern associated with the linear-propagation effective apodization transmit case were found to be in good agreement with the detailed spatial characteristics of the nonlinearly generated second harmonic field. The effects of the abdominal wall tissue aberrators are apparent for all three of the beam profiles studied. Consistent with the improved image quality associated with harmonic imaging, the aberrated nonlinearly generated second harmonic beam was shown to remain more compact than the corresponding aberrated fundamental beam patterns in the presence of the interposed aberrator
Beschreibung:Date Completed 05.09.2006
Date Revised 17.09.2019
published: Print
Citation Status MEDLINE
ISSN:0885-3010