Set-up and experimental parameters of the protein crystallography beamline at the Brazilian National Synchrotron Laboratory

The Brazilian National Synchrotron Light Laboratory (LNLS) has a dedicated protein crystallography beamline. The optical elements of the beamline include an elastically bent cylindrical mirror and a triangular bent-crystal monochromator, which focus synchrotron radiation at the position of the sampl...

Ausführliche Beschreibung

Bibliographische Detailangaben
Veröffentlicht in:Journal of synchrotron radiation. - 1994. - 5(1998), Pt 2 vom: 01. März, Seite 72-6
1. Verfasser: Polikarpov, I (VerfasserIn)
Weitere Verfasser: Perles, L A, de Oliveira, R T, Oliva, G, Castellano, E E, Garratt, R C, Craievich, A
Format: Aufsatz
Sprache:English
Veröffentlicht: 1998
Zugriff auf das übergeordnete Werk:Journal of synchrotron radiation
Schlagworte:Journal Article
LEADER 01000naa a22002652 4500
001 NLM162696698
003 DE-627
005 20231223094753.0
007 tu
008 231223s1998 xx ||||| 00| ||eng c
028 5 2 |a pubmed24n0542.xml 
035 |a (DE-627)NLM162696698 
035 |a (NLM)16687806 
040 |a DE-627  |b ger  |c DE-627  |e rakwb 
041 |a eng 
100 1 |a Polikarpov, I  |e verfasserin  |4 aut 
245 1 0 |a Set-up and experimental parameters of the protein crystallography beamline at the Brazilian National Synchrotron Laboratory 
264 1 |c 1998 
336 |a Text  |b txt  |2 rdacontent 
337 |a ohne Hilfsmittel zu benutzen  |b n  |2 rdamedia 
338 |a Band  |b nc  |2 rdacarrier 
500 |a Date Completed 24.04.2008 
500 |a Date Revised 11.05.2006 
500 |a published: Print 
500 |a Citation Status PubMed-not-MEDLINE 
520 |a The Brazilian National Synchrotron Light Laboratory (LNLS) has a dedicated protein crystallography beamline. The optical elements of the beamline include an elastically bent cylindrical mirror and a triangular bent-crystal monochromator, which focus synchrotron radiation at the position of the sample in the vertical and horizontal planes, respectively. The monochromatic radiation is tunable between 2.0 and 1.2 A with the optimum wavelengths from 1.3 to 1.6 A, chosen to maximize the photon flux from the bending magnets of the storage ring (1.37 GeV). Diffraction images are recorded on a 345 mm-diameter MarResearch image-plate detector system with on-line readout. The experimental parameters of the beamline, such as the integral monochromatic flux and focus size, have been measured. The size of the beam at the position of the focal point is 0.5 x 0.5 mm(2). The flux density is between 4.4 x 10(10) and 8 x 10(10) photons s(-1) mm(-2) for wavelengths from 1.28 to 1.6 A. The energy resolution is sufficient to measure absorption edges of elements between 1.28 and 2 A. The facility, intended to serve the national and international community, has been commissioned and is available for users 
650 4 |a Journal Article 
700 1 |a Perles, L A  |e verfasserin  |4 aut 
700 1 |a de Oliveira, R T  |e verfasserin  |4 aut 
700 1 |a Oliva, G  |e verfasserin  |4 aut 
700 1 |a Castellano, E E  |e verfasserin  |4 aut 
700 1 |a Garratt, R C  |e verfasserin  |4 aut 
700 1 |a Craievich, A  |e verfasserin  |4 aut 
773 0 8 |i Enthalten in  |t Journal of synchrotron radiation  |d 1994  |g 5(1998), Pt 2 vom: 01. März, Seite 72-6  |w (DE-627)NLM09824129X  |x 1600-5775  |7 nnns 
773 1 8 |g volume:5  |g year:1998  |g number:Pt 2  |g day:01  |g month:03  |g pages:72-6 
912 |a GBV_USEFLAG_A 
912 |a SYSFLAG_A 
912 |a GBV_NLM 
912 |a GBV_ILN_40 
912 |a GBV_ILN_350 
912 |a GBV_ILN_2005 
951 |a AR 
952 |d 5  |j 1998  |e Pt 2  |b 01  |c 03  |h 72-6