X-ray diffractometer combining synchrotron radiation and pulsed magnetic fields up to 40 T
A synchrotron X-ray diffractometer incorporating a pulsed field magnet for high fields up to 40 T has been developed and a detailed description of this instrument is reported. The pulsed field magnet is composed of two coaxial coils with a gap of 3 mm at the mid-plane for passage of the X-rays. The...
Veröffentlicht in: | Journal of synchrotron radiation. - 1994. - 13(2006), Pt 3 vom: 15. Mai, Seite 271-4 |
---|---|
1. Verfasser: | |
Weitere Verfasser: | , , , , , , , , , , , |
Format: | Aufsatz |
Sprache: | English |
Veröffentlicht: |
2006
|
Zugriff auf das übergeordnete Werk: | Journal of synchrotron radiation |
Schlagworte: | Journal Article |
Zusammenfassung: | A synchrotron X-ray diffractometer incorporating a pulsed field magnet for high fields up to 40 T has been developed and a detailed description of this instrument is reported. The pulsed field magnet is composed of two coaxial coils with a gap of 3 mm at the mid-plane for passage of the X-rays. The pixel detector PILATUS 100K is used to store the diffracted X-rays. As a test of this instrument, X-ray diffraction by a powder sample of the antiferromagnet CoO is measured below the Néel temperature. A field-dependent lattice distortion of CoO due to magnetostriction is observed up to 38 T |
---|---|
Beschreibung: | Date Completed 20.06.2006 Date Revised 28.04.2006 published: Print-Electronic Citation Status PubMed-not-MEDLINE |