A multiwire proportional counter for very high counting rates
Preliminary measurements in a proportional counter with two independently counting wires showed that counting rates up to 10(6) counts s(-1) wire(-1) can be reached without critical loss in the 'true versus measured' linearity relation. Results obtained with a detector containing 30 active...
Veröffentlicht in: | Journal of synchrotron radiation. - 1998. - 5(1998), Pt 3 vom: 01. Mai, Seite 860-2 |
---|---|
1. Verfasser: | |
Weitere Verfasser: | , , , |
Format: | Aufsatz |
Sprache: | English |
Veröffentlicht: |
1998
|
Zugriff auf das übergeordnete Werk: | Journal of synchrotron radiation |
Schlagworte: | Journal Article |
Zusammenfassung: | Preliminary measurements in a proportional counter with two independently counting wires showed that counting rates up to 10(6) counts s(-1) wire(-1) can be reached without critical loss in the 'true versus measured' linearity relation. Results obtained with a detector containing 30 active wires (2 mm pitch) are presented. With each wire is associated a fast pre-amplifier and a discriminator channel. Global counting rates in excess of 10(7) events s(-1) are reported. Dead-time losses are corrected by use of simple mathematical-modelling functions. Data-acquisition systems are described for one-dimensional (real-time) and two-dimensional (off-line) position-sensitive detection systems |
---|---|
Beschreibung: | Date Completed 02.10.2012 Date Revised 20.07.2004 published: Print-Electronic Citation Status PubMed-not-MEDLINE |
ISSN: | 0909-0495 |