Preliminary measurements in a proportional counter with two independently counting wires showed that counting rates up to 10(6) counts s(-1) wire(-1) can be reached without critical loss in the 'true versus measured' linearity relation. Results obtained with a detector containing 30 active...

Ausführliche Beschreibung

Bibliographische Detailangaben
Veröffentlicht in:Journal of synchrotron radiation. - 1998. - 5(1998), Pt 3 vom: 01. Mai, Seite 860-2
1. Verfasser: Barbosa, A F (VerfasserIn)
Weitere Verfasser: Guedes, G P, Tamura, E, Pepe, I M, Oliveira, N B
Format: Aufsatz
Sprache:English
Veröffentlicht: 1998
Zugriff auf das übergeordnete Werk:Journal of synchrotron radiation
Schlagworte:Journal Article
Beschreibung
Zusammenfassung:Preliminary measurements in a proportional counter with two independently counting wires showed that counting rates up to 10(6) counts s(-1) wire(-1) can be reached without critical loss in the 'true versus measured' linearity relation. Results obtained with a detector containing 30 active wires (2 mm pitch) are presented. With each wire is associated a fast pre-amplifier and a discriminator channel. Global counting rates in excess of 10(7) events s(-1) are reported. Dead-time losses are corrected by use of simple mathematical-modelling functions. Data-acquisition systems are described for one-dimensional (real-time) and two-dimensional (off-line) position-sensitive detection systems