Electronics for SPring-8 X-ray beam monitors
A sensitive current-measuring system is required to construct a highly sensitive X-ray beam-position monitor (XBPM). A current-voltage converter (I/V) which can measure currents between 0.1 nA and 10 mA was designed, and the signal processing system of the XBPM was constucted using this I/V. This sy...
Veröffentlicht in: | Journal of synchrotron radiation. - 1994. - 5(1998), Pt 3 vom: 01. Mai, Seite 630-1 |
---|---|
1. Verfasser: | |
Weitere Verfasser: | , , , |
Format: | Aufsatz |
Sprache: | English |
Veröffentlicht: |
1998
|
Zugriff auf das übergeordnete Werk: | Journal of synchrotron radiation |
Schlagworte: | Journal Article |
Zusammenfassung: | A sensitive current-measuring system is required to construct a highly sensitive X-ray beam-position monitor (XBPM). A current-voltage converter (I/V) which can measure currents between 0.1 nA and 10 mA was designed, and the signal processing system of the XBPM was constucted using this I/V. This system was used in beamline commissioning. Beam-position data standard deviations of sigma approximately 3 micro m for the bending-magnet beamline, and sigma(x) approximately 3 micro m and sigma(y) approximately 1 micro m for the insertion-device beamline were obtained during the beamline commissioning |
---|---|
Beschreibung: | Date Completed 02.10.2012 Date Revised 20.07.2004 published: Print-Electronic Citation Status PubMed-not-MEDLINE |