Development of a three-electrode-lens drift tube for time-of-flight mass spectrometry
A three-electrode-lens drift tube for time-of-flight mass spectrometry (TOF-MS) has been developed for utilizing a detector to observe photon-stimulated desorption (PSD). In spite of a small detection area, the detector has a high detection efficiency and durability to reactive gas atmosphere at hig...
Veröffentlicht in: | Journal of synchrotron radiation. - 1994. - 5(1998), Pt 3 vom: 01. Mai, Seite 612-4 |
---|---|
1. Verfasser: | |
Weitere Verfasser: | , , , , |
Format: | Aufsatz |
Sprache: | English |
Veröffentlicht: |
1998
|
Zugriff auf das übergeordnete Werk: | Journal of synchrotron radiation |
Schlagworte: | Journal Article |
Zusammenfassung: | A three-electrode-lens drift tube for time-of-flight mass spectrometry (TOF-MS) has been developed for utilizing a detector to observe photon-stimulated desorption (PSD). In spite of a small detection area, the detector has a high detection efficiency and durability to reactive gas atmosphere at high pressure. The TOF-MS performance of the drift tube was examined for PSD using single-bunch-mode synchrotron radiation on a dichlorosilane (SiH(2)Cl(2))-saturated Si(001) surface. The measured acceleration and focusing-voltage dependences of the time of flight, intensity and full width at half-maximum for the peak of H(+) and Cl(+) PSD ions are discussed in terms of the numerical calculations of ion trajectories and focusing characteristic of the drift tube |
---|---|
Beschreibung: | Date Completed 02.10.2012 Date Revised 20.07.2004 published: Print-Electronic Citation Status PubMed-not-MEDLINE |