Design of a high-flux and high-resolution VUV bending-magnet beamline

A high-flux and high-resolution VUV beamline (4-40 eV) has been designed and is under construction at SRRC. This beamline, which collects 50 mrad of horizontal radiation, uses a 6 m cylindrical-grating monochromator with an incident angle of 70 degrees instead of the conventional normal-incidence-mo...

Ausführliche Beschreibung

Bibliographische Detailangaben
Veröffentlicht in:Journal of synchrotron radiation. - 1998. - 5(1998), Pt 3 vom: 01. Mai, Seite 562-4
1. Verfasser: Hsieh, T F (VerfasserIn)
Weitere Verfasser: Huang, L R, Chung, S C, Dann, T E, Tseng, P C, Chen, C T, Tsang, K L
Format: Aufsatz
Sprache:English
Veröffentlicht: 1998
Zugriff auf das übergeordnete Werk:Journal of synchrotron radiation
Schlagworte:Journal Article
LEADER 01000caa a22002652 4500
001 NLM149469985
003 DE-627
005 20250205170015.0
007 tu
008 231223s1998 xx ||||| 00| ||eng c
028 5 2 |a pubmed25n0498.xml 
035 |a (DE-627)NLM149469985 
035 |a (NLM)15263579 
040 |a DE-627  |b ger  |c DE-627  |e rakwb 
041 |a eng 
100 1 |a Hsieh, T F  |e verfasserin  |4 aut 
245 1 0 |a Design of a high-flux and high-resolution VUV bending-magnet beamline 
264 1 |c 1998 
336 |a Text  |b txt  |2 rdacontent 
337 |a ohne Hilfsmittel zu benutzen  |b n  |2 rdamedia 
338 |a Band  |b nc  |2 rdacarrier 
500 |a Date Completed 02.10.2012 
500 |a Date Revised 20.07.2004 
500 |a published: Print-Electronic 
500 |a Citation Status PubMed-not-MEDLINE 
520 |a A high-flux and high-resolution VUV beamline (4-40 eV) has been designed and is under construction at SRRC. This beamline, which collects 50 mrad of horizontal radiation, uses a 6 m cylindrical-grating monochromator with an incident angle of 70 degrees instead of the conventional normal-incidence-monochromator (NIM) design. Special features, such as movable entrance slit, bendable vertical focusing mirror and movable curved exit slit, are employed to enhance greatly the beamline performance. With both slit openings set at 10 micro m, the energy-resolving power can reach as high as 70000. Photon fluxes of 1 x 10(13) and 1 x 10(10) photons s(-1) are calculated for energy-resolving powers of 1000 and 40000, respectively. The best image size at the sample position is smaller than 0.45 x 0.2 mm 
650 4 |a Journal Article 
700 1 |a Huang, L R  |e verfasserin  |4 aut 
700 1 |a Chung, S C  |e verfasserin  |4 aut 
700 1 |a Dann, T E  |e verfasserin  |4 aut 
700 1 |a Tseng, P C  |e verfasserin  |4 aut 
700 1 |a Chen, C T  |e verfasserin  |4 aut 
700 1 |a Tsang, K L  |e verfasserin  |4 aut 
773 0 8 |i Enthalten in  |t Journal of synchrotron radiation  |d 1998  |g 5(1998), Pt 3 vom: 01. Mai, Seite 562-4  |w (DE-627)NLM09824129X  |x 0909-0495  |7 nnns 
773 1 8 |g volume:5  |g year:1998  |g number:Pt 3  |g day:01  |g month:05  |g pages:562-4 
912 |a GBV_USEFLAG_A 
912 |a SYSFLAG_A 
912 |a GBV_NLM 
912 |a GBV_ILN_40 
912 |a GBV_ILN_350 
912 |a GBV_ILN_2005 
951 |a AR 
952 |d 5  |j 1998  |e Pt 3  |b 01  |c 05  |h 562-4