On the thermal effects associated with radiation force imaging of soft tissue

Several laboratories are investigating the use of acoustic radiation force to image the mechanical properties of tissue. Acoustic Radiation Force Impulse (ARFI) imaging is one approach that uses brief, high-intensity, focused ultrasound pulses to generate radiation force in tissue. This radiation fo...

Description complète

Détails bibliographiques
Publié dans:IEEE transactions on ultrasonics, ferroelectrics, and frequency control. - 1986. - 51(2004), 5 vom: 19. Mai, Seite 551-65
Auteur principal: Palmeri, Mark L (Auteur)
Autres auteurs: Nightingale, Kathryn R
Format: Article
Langue:English
Publié: 2004
Accès à la collection:IEEE transactions on ultrasonics, ferroelectrics, and frequency control
Sujets:Comparative Study Evaluation Study Journal Article Research Support, Non-U.S. Gov't Research Support, U.S. Gov't, P.H.S.
Description
Résumé:Several laboratories are investigating the use of acoustic radiation force to image the mechanical properties of tissue. Acoustic Radiation Force Impulse (ARFI) imaging is one approach that uses brief, high-intensity, focused ultrasound pulses to generate radiation force in tissue. This radiation force generates tissue displacements that are tracked using conventional correlation-based ultrasound methods. The tissue response provides a mechanism to discern mechanical properties of the tissue. The acoustic energy that is absorbed by tissue generates radiation force and tissue heating. A finite element methods model of acoustic heating has been developed that models the thermal response of different tissues during short duration radiation force application. The beam sequences and focal configurations used during ARFI imaging are modeled herein; the results of these thermal models can be extended to the heating due to absorption associated with other radiation force-based imaging modalities. ARFI-induced thermal diffusivity patterns are functions of the transducer f-number, the tissue absorption, and the temporal and spatial spacing of adjacent ARFI interrogations. Cooling time constants are on the order of several seconds. Tissue displacement due to thermal expansion is negligible for ARFI imaging. Changes in sound speed due to temperature changes can be appreciable. These thermal models demonstrate that ARFI imaging of soft tissue is safe, although thermal response must be monitored when ARFI beam sequences are being developed
Description:Date Completed 23.07.2004
Date Revised 10.12.2019
published: Print
Citation Status MEDLINE
ISSN:1525-8955