A new neutron single-crystal diffractometer dedicated for biological macromolecules (BIX-4)

A new neutron single-crystal diffractometer (BIX-4) has been constructed at 1G-B port of JRR-3M in JAERI. Since at 1G-B port another diffractometer for biology, BIX-3, and a high-resolution powder diffractometer (HRPD) coexist, the monochromator house needed to be reconstructed. The main architectur...

Ausführliche Beschreibung

Bibliographische Detailangaben
Veröffentlicht in:Journal of synchrotron radiation. - 1994. - 11(2004), Pt 1 vom: 01. Jan., Seite 68-71
1. Verfasser: Kurihara, Kazuo (VerfasserIn)
Weitere Verfasser: Tanaka, Ichiro, Niimura, Nobuo, Refai Muslih, Muhammad, Ostermann, Andreas
Format: Aufsatz
Sprache:English
Veröffentlicht: 2004
Zugriff auf das übergeordnete Werk:Journal of synchrotron radiation
Schlagworte:Evaluation Study Journal Article Research Support, Non-U.S. Gov't Biopolymers Macromolecular Substances Proteins
LEADER 01000naa a22002652 4500
001 NLM143615750
003 DE-627
005 20231223031302.0
007 tu
008 231223s2004 xx ||||| 00| ||eng c
028 5 2 |a pubmed24n0479.xml 
035 |a (DE-627)NLM143615750 
035 |a (NLM)14646137 
040 |a DE-627  |b ger  |c DE-627  |e rakwb 
041 |a eng 
100 1 |a Kurihara, Kazuo  |e verfasserin  |4 aut 
245 1 2 |a A new neutron single-crystal diffractometer dedicated for biological macromolecules (BIX-4) 
264 1 |c 2004 
336 |a Text  |b txt  |2 rdacontent 
337 |a ohne Hilfsmittel zu benutzen  |b n  |2 rdamedia 
338 |a Band  |b nc  |2 rdacarrier 
500 |a Date Completed 15.04.2004 
500 |a Date Revised 10.12.2019 
500 |a published: Print-Electronic 
500 |a Citation Status MEDLINE 
520 |a A new neutron single-crystal diffractometer (BIX-4) has been constructed at 1G-B port of JRR-3M in JAERI. Since at 1G-B port another diffractometer for biology, BIX-3, and a high-resolution powder diffractometer (HRPD) coexist, the monochromator house needed to be reconstructed. The main architecture of BIX-4 is based on that of BIX-3. BIX-4 uses an elastically-bent perfect-Si crystal monochromator and neutron imaging plates as BIX-3. In addition, several optimizations of the monochromator and modifications from previous BIX-3 are carried out. The final gain of the neutron intensity at the detector position is estimated to be 2.5 times larger than previous BIX-3. That higher performance increases the opportunities to apply neutron crystallography to biological macromolecules which give only weak reflections and/or small crystals 
650 4 |a Evaluation Study 
650 4 |a Journal Article 
650 4 |a Research Support, Non-U.S. Gov't 
650 7 |a Biopolymers  |2 NLM 
650 7 |a Macromolecular Substances  |2 NLM 
650 7 |a Proteins  |2 NLM 
700 1 |a Tanaka, Ichiro  |e verfasserin  |4 aut 
700 1 |a Niimura, Nobuo  |e verfasserin  |4 aut 
700 1 |a Refai Muslih, Muhammad  |e verfasserin  |4 aut 
700 1 |a Ostermann, Andreas  |e verfasserin  |4 aut 
773 0 8 |i Enthalten in  |t Journal of synchrotron radiation  |d 1994  |g 11(2004), Pt 1 vom: 01. Jan., Seite 68-71  |w (DE-627)NLM09824129X  |x 1600-5775  |7 nnns 
773 1 8 |g volume:11  |g year:2004  |g number:Pt 1  |g day:01  |g month:01  |g pages:68-71 
912 |a GBV_USEFLAG_A 
912 |a SYSFLAG_A 
912 |a GBV_NLM 
912 |a GBV_ILN_40 
912 |a GBV_ILN_350 
912 |a GBV_ILN_2005 
951 |a AR 
952 |d 11  |j 2004  |e Pt 1  |b 01  |c 01  |h 68-71