This work develops a class of ultrasound phase aberration correction/autofocusing algorithms that are based upon the properties of the covariance matrix of the channel signals for time-delay focused resolution/speckle cells. The scaled covariance matrix (SCM) algorithms are designed to blindly estim...

Ausführliche Beschreibung

Bibliographische Detailangaben
Veröffentlicht in:IEEE transactions on ultrasonics, ferroelectrics, and frequency control. - 1999. - 50(2003), 7 vom: 26. Juli, Seite 795-804
1. Verfasser: Silverstein, Seth D (VerfasserIn)
Weitere Verfasser: Ceperley, Daniel P
Format: Aufsatz
Sprache:English
Veröffentlicht: 2003
Zugriff auf das übergeordnete Werk:IEEE transactions on ultrasonics, ferroelectrics, and frequency control
Schlagworte:Comparative Study Evaluation Study Journal Article Research Support, U.S. Gov't, Non-P.H.S. Validation Study
LEADER 01000caa a22002652 4500
001 NLM12651156X
003 DE-627
005 20250204050445.0
007 tu
008 231222s2003 xx ||||| 00| ||eng c
028 5 2 |a pubmed25n0422.xml 
035 |a (DE-627)NLM12651156X 
035 |a (NLM)12894914 
040 |a DE-627  |b ger  |c DE-627  |e rakwb 
041 |a eng 
100 1 |a Silverstein, Seth D  |e verfasserin  |4 aut 
245 1 0 |a Autofocusing in medical ultrasound  |b the scaled covariance matrix algorithm 
264 1 |c 2003 
336 |a Text  |b txt  |2 rdacontent 
337 |a ohne Hilfsmittel zu benutzen  |b n  |2 rdamedia 
338 |a Band  |b nc  |2 rdacarrier 
500 |a Date Completed 04.09.2003 
500 |a Date Revised 10.12.2019 
500 |a published: Print 
500 |a Citation Status MEDLINE 
520 |a This work develops a class of ultrasound phase aberration correction/autofocusing algorithms that are based upon the properties of the covariance matrix of the channel signals for time-delay focused resolution/speckle cells. The scaled covariance matrix (SCM) algorithms are designed to blindly estimate and correct focusing timing errors due to thin layers of unanticipated fatty tissue located in the near field of the transducer array. An important aspect of the algorithm is that the scaling of the covariance matrix elements fundamentally establishes a channel independent phase reference relative to which the aberrant channel phases are estimated. The model development involved the combination of a rigorous mathematical analysis of the scattering of ultrasound in random scattering media and extensive statistical simulation studies with phase aberrations imposed upon both the transmit and received channel signals. Under the assumption of a near field aberration model, the statistical simulation analyses showed that the SCM algorithms in simulation are capable of accurately estimating relative time delay channel errors with rms timing errors up to approximately 62 ns, with interchannel correlation lengths as short as 1.4 mm 
650 4 |a Comparative Study 
650 4 |a Evaluation Study 
650 4 |a Journal Article 
650 4 |a Research Support, U.S. Gov't, Non-P.H.S. 
650 4 |a Validation Study 
700 1 |a Ceperley, Daniel P  |e verfasserin  |4 aut 
773 0 8 |i Enthalten in  |t IEEE transactions on ultrasonics, ferroelectrics, and frequency control  |d 1999  |g 50(2003), 7 vom: 26. Juli, Seite 795-804  |w (DE-627)NLM098181017  |x 0885-3010  |7 nnns 
773 1 8 |g volume:50  |g year:2003  |g number:7  |g day:26  |g month:07  |g pages:795-804 
912 |a GBV_USEFLAG_A 
912 |a SYSFLAG_A 
912 |a GBV_NLM 
912 |a GBV_ILN_22 
912 |a GBV_ILN_24 
912 |a GBV_ILN_350 
951 |a AR 
952 |d 50  |j 2003  |e 7  |b 26  |c 07  |h 795-804