Fixed-height exit bender of synchrotron X-rays above 40 keV

A crystal bender for sagittal focusing has been designed for standard monochromators at SPring-8. The bender does not move the position of the crystal center when the bending radius is changed. Sagittal focusing from 40 keV to 60 keV was achieved by using Si(311) double crystals. The flux density of...

Ausführliche Beschreibung

Bibliographische Detailangaben
Veröffentlicht in:Journal of synchrotron radiation. - 1994. - 8(2001), 1 vom: 01. Jan., Seite 18-21
1. Verfasser: Yoneda, Y (VerfasserIn)
Weitere Verfasser: Matsumoto, N, Furukawa, Y, Ishikawa, T
Format: Aufsatz
Sprache:English
Veröffentlicht: 2001
Zugriff auf das übergeordnete Werk:Journal of synchrotron radiation
Schlagworte:Journal Article
Beschreibung
Zusammenfassung:A crystal bender for sagittal focusing has been designed for standard monochromators at SPring-8. The bender does not move the position of the crystal center when the bending radius is changed. Sagittal focusing from 40 keV to 60 keV was achieved by using Si(311) double crystals. The flux density of the focused beam measured at 40 keV was 15 times higher than that of the unfocused beam. The height deviation of the focused beam throughout the measured energy range was within +/- 0.15 mm
Beschreibung:Date Completed 23.08.2001
Date Revised 05.06.2019
published: Print
Citation Status MEDLINE
ISSN:1600-5775