Evaluation of both local exhaust and cycle purges to reduce operator exposure and environmental levels of ethylene oxide

A local exhaust manifold was attached to two different tabletop ethylene oxide sterilizers. One tabletop sterilizer had a built-in cycle purge feature, while the other had the standard single terminal vacuum cycle. Operator exposure and environmental levels of ethylene oxide were determined with and...

Ausführliche Beschreibung

Bibliographische Detailangaben
Veröffentlicht in:Sterile world. - 1981. - 5(1983), 2 vom: 01. Apr., Seite 2-4
1. Verfasser: Samuels, T M (VerfasserIn)
Format: Aufsatz
Sprache:English
Veröffentlicht: 1983
Zugriff auf das übergeordnete Werk:Sterile world
Schlagworte:Journal Article Air Pollutants Air Pollutants, Occupational Ethylene Oxide JJH7GNN18P
Beschreibung
Zusammenfassung:A local exhaust manifold was attached to two different tabletop ethylene oxide sterilizers. One tabletop sterilizer had a built-in cycle purge feature, while the other had the standard single terminal vacuum cycle. Operator exposure and environmental levels of ethylene oxide were determined with and without local exhaust, and with and without purge cycle. Results from both charcoal tube collection samples and infra-red analysis clearly demonstrated the benefit of a local exhaust coupled with cycle purge in reducing operator exposure to and environmental contamination by ethylene oxide to nearly undetectable levels
Beschreibung:Date Completed 11.08.1983
Date Revised 21.11.2013
published: Print
Citation Status MEDLINE