The effect of disinfecting procedure on the glucose concentration using a personal glucose meter

• Disinfection cause lowering of the measurement. • After disinfection, the best option is to wait 30 s before measurement. • Thinner lancet after disinfection cause greater lowering of the results.

Bibliographische Detailangaben
Veröffentlicht in:A pilot randomized controlled trial examining the feasibility, acceptability, and efficacy of Adapted Motivational Interviewing for post-operative bariatric surgery patients. - 2016. - Amsterdam [u.a.]
1. Verfasser: Jońca, Mateusz (VerfasserIn)
Weitere Verfasser: Krótki, Filipina (BerichterstatterIn), Tomasik, Przemysław (BerichterstatterIn)
Format: Online-Aufsatz
Sprache:English
Veröffentlicht: 2021
Zugriff auf das übergeordnete Werk:A pilot randomized controlled trial examining the feasibility, acceptability, and efficacy of Adapted Motivational Interviewing for post-operative bariatric surgery patients
Schlagworte:Glucose concentration POCT Disinfectant Personal glucose meters
Umfang:5