Combined X-ray and neutron single-crystal diffraction in diamond anvil cells
© Andrzej Grzechnik et al. 2020.
| Publié dans: | Journal of applied crystallography. - 1998. - 53(2020), Pt 1 vom: 01. Feb., Seite 9-14 |
|---|---|
| Auteur principal: | |
| Autres auteurs: | , , |
| Format: | Article en ligne |
| Langue: | English |
| Publié: |
2020
|
| Accès à la collection: | Journal of applied crystallography |
| Sujets: | Journal Article diamond anvil cells high pressure single-crystal X-ray diffraction single-crystal neutron diffraction |
| Résumé: | © Andrzej Grzechnik et al. 2020. It is shown that it is possible to perform combined X-ray and neutron single-crystal studies in the same diamond anvil cell (DAC). A modified Merrill-Bassett DAC equipped with an inflatable membrane filled with He gas has been developed. It can be used on laboratory X-ray and synchrotron diffractometers as well as on neutron instruments. The data processing procedures and a joint structural refinement of the high-pressure synchrotron and neutron single-crystal data are presented and discussed for the first time |
|---|---|
| Description: | Date Revised 28.03.2024 published: Electronic-eCollection Citation Status PubMed-not-MEDLINE |
| ISSN: | 0021-8898 |
| DOI: | 10.1107/S1600576719014201 |