Scleral Transillumination With Digital Heads-Up Display : A Novel Technique for Visualization During Vitrectomy Surgery

Copyright 2018, SLACK Incorporated.

Bibliographische Detailangaben
Veröffentlicht in:Ophthalmic surgery, lasers & imaging retina. - 2013. - 49(2018), 6 vom: 01. Juni, Seite 436-439
1. Verfasser: Todorich, Bozho (VerfasserIn)
Weitere Verfasser: Stem, Maxwell S, Hassan, Tarek S, Williams, George A, Faia, Lisa J
Format: Online-Aufsatz
Sprache:English
Veröffentlicht: 2018
Zugriff auf das übergeordnete Werk:Ophthalmic surgery, lasers & imaging retina
Schlagworte:Journal Article
Beschreibung
Zusammenfassung:Copyright 2018, SLACK Incorporated.
BACKGROUND AND OBJECTIVE: To describe a novel technique of scleral indentation and transillumination for single-surgeon, unassisted vitrectomy and vitreous base shaving enhanced with a digital heads-up display system (NGENUITY 3D Visualization System; Alcon, Fort Worth, TX)
PATIENTS AND METHODS: This technique was utilized in six eyes of six patients during vitrectomy surgery for common vitreoretinal surgical diagnoses. In each case, the transillumination was performed with the traditional intraocular light pipe set at 100% power, placed obliquely just posterior to the vitreous base insertion, with or without a transillumination adapter. The visualization of the vitreous cavity was digitally enhanced using a heads-up display system (NGENUITY 3D) with light amplification settings increased to near-maximal gain. In each case, the adequacy of the surgical view was judged intraoperatively by two independent surgeons who shared the same surgical view as the primary surgeon
RESULTS: In this series, the surgical view provided by the scleral transillumination was deemed adequate to safely perform surgery in five of six cases. In the one patient in whom this was not the case, vitrectomy was completed using traditional endo-illumination and scleral depression performed by a skilled assistant. Lighter fundus pigmentation, myopia, thin sclera, and absence of dense peripheral media opacities were associated with improved view with scleral transillumination. There were no intraoperative complications
CONCLUSION: Digitally enhanced scleral transillumination affords surgeons another option for safe and effective simultaneous scleral depression and illumination for unassisted peripheral vitrectomy. [Ophthalmic Surg Lasers Imaging Retina. 2018;49:436-439.]
Beschreibung:Date Completed 14.03.2019
Date Revised 14.03.2019
published: Print
Citation Status MEDLINE
ISSN:2325-8179
DOI:10.3928/23258160-20180601-08