Toward a reference ultrasonic cavitation vessel : Part 2--investigating the spatial variation and acoustic pressure threshold of inertial cavitation in a 25 kHz ultrasound field
As part of an ongoing project to establish a reference facility for acoustic cavitation at the National Physical Laboratory (NPL), carefully controlled studies on a 25 kHz, 1.8 kW cylindrical vessel are described. Using a patented high-frequency acoustic emission detection method and a sonar hydroph...
Ausführliche Beschreibung